Hemoperfusion machine JF-800A is a drive machine for blood extracorporeal circulation with automatic anticoagulant-injection device, blood warming device, safety monitoring device for pressure, liquid level & air bubble. It is widely used in blood purification treatment & other extracorporeal circulation treatment, such as hemoperfusion.
Specifications
Exterior Dimensions: 310W*240D*430H mm (excluding the medical trolley )
Weight: 15 kg (excluding the medical trolley )
Power supply: AC220V+/-10% 50H+/-2%
Power: about 150VA
Fuse: Ø5*20 mm, T2AL250VAC
Electroshock protection level: Class A type B
Suitable tube: inner diameter Ø8 mm, Ø6 mm
Target volume range: 1~99999 ml
Flux of blood pump: 9~450ml/min(Ø8), 6~300ml/min(Ø6)
Precision of blood pump flux: +/-5%
Range of heparin pump flux: -20~+38 KPa (1KPa)
Precision of heparin pump flux: +/-5%
Display range of the arterial pressure: -20~+38 KPa (1KPa)
Range of the arterial warning: -20~+38 KPa (1KPa)
Display range of the venous pressure: -18~+40KPa KPa (1KPa)
Range of the venous warning: -18~+40KPa (1KPa)
Range of the warming mediator: 36~41℃(±0.5℃)
|